Single Channel, Dual Channel, 4 Channel RF Lesion Generator Radiofrequency Thermal Coagulation Ablation System

Product Details
Customization: Available
Classification: Pain Medicine
Type: RF Lesion Generator
Gold Member Since 2024

Suppliers with verified business licenses

Audited Supplier

Audited by an independent third-party inspection agency

to see all verified strength labels (11)
  • Single Channel, Dual Channel, 4 Channel RF Lesion Generator Radiofrequency Thermal Coagulation Ablation System
  • Single Channel, Dual Channel, 4 Channel RF Lesion Generator Radiofrequency Thermal Coagulation Ablation System
  • Single Channel, Dual Channel, 4 Channel RF Lesion Generator Radiofrequency Thermal Coagulation Ablation System
  • Single Channel, Dual Channel, 4 Channel RF Lesion Generator Radiofrequency Thermal Coagulation Ablation System
Find Similar Products

Basic Info.

Model NO.
SL-LPS400K
Certification
Equipment Factory Inspection Report
Group
People with Nerve Pain
Transport Package
Carton Packaging
Specification
Spine, nerves
Origin
China
Production Capacity
1000 Set/Month

Product Description

Single Channel, Dual Channel, 4 Channel RF Lesion Generator Radiofrequency Thermal Coagulation Ablation SystemRF thermal coagulation is to emit a high frequency RF current through the magnetic field. The RF current produces a changing magnetic field at the tip of the working electrode, so that the molecular movement in the target tissue covered by the magnetic field friction generates heat, and thermal coagulation damages the organization of the target area. It is an international advanced minimally invasive technology and the world's cutting-edge pain treatment method, with the characteristics of minimally invasive, qualitative, quantitative, definite efficacy and less complications.

2 Working principle

The RF thermal coagulation function of Thermo-coagulation Unit uses a disposable adhesive plate as the neutral electrode, and the thermal coagulation electrode transmits the RF thermal coagulation energy through the human tissue and the neutral electrode to produce the thermal coagulation treatment effect.
Single Channel, Dual Channel, 4 Channel RF Lesion Generator Radiofrequency Thermal Coagulation Ablation System
(2) Audio output table
order number Audio properties signalment remarks
1 Normal buzzer 1 Bbass for 0.5 seconds Normal initialization
2 Fault sound Low pitch tone two tones: drops Neutral electrode failure
3 Confirm sound High and low tones of two tones Neutral electrode fault recovery
4 The RFA output is effectively effectively High pitch continuous sound During the RFA effective output
5 RF coagulation output is effectively effectively Bass continuous sound During the effective output of RF coagulation
6 Fault sound The same tone of two sounds: drops With ault fault, see Error Display
(3) Handling with common problems
phenomenon cause  The exclusion method
Turn on the power switch, and the device does not respond. 1, the network power outage;
2. The power cord is not well connected;
3. Ffuse.
1. Connect to the power supply;
2. Connect the power cord;
3. Replace the fuse.
During the normal working setting, the output power intensity setting is significantly reduced or the equipment does not work normally. 1. Poor contact of neutral electrodes;
2. The use method is incorrect.
1. Connect the neutral electrode correctly;
2. Repair or replace the accessories;
3, ask the manufacturer to solve.


 