order number | Audio properties | signalment | remarks |
1 | Normal buzzer 1 | Bbass for 0.5 seconds | Normal initialization |
2 | Fault sound | Low pitch tone two tones: drops | Neutral electrode failure |
3 | Confirm sound | High and low tones of two tones | Neutral electrode fault recovery |
4 | The RFA output is effectively effectively | High pitch continuous sound | During the RFA effective output |
5 | RF coagulation output is effectively effectively | Bass continuous sound | During the effective output of RF coagulation |
6 | Fault sound | The same tone of two sounds: drops | With ault fault, see Error Display |
phenomenon | cause | The exclusion method |
Turn on the power switch, and the device does not respond. | 1, the network power outage; 2. The power cord is not well connected; 3. Ffuse. |
1. Connect to the power supply; 2. Connect the power cord; 3. Replace the fuse. |
During the normal working setting, the output power intensity setting is significantly reduced or the equipment does not work normally. | 1. Poor contact of neutral electrodes; 2. The use method is incorrect. |
1. Connect the neutral electrode correctly; 2. Repair or replace the accessories; 3, ask the manufacturer to solve. |